FDA Adverse Event
Malfunction
Summary report: N
OT PING ENHANCED METER
MDR report key: 3962517
·
Received July 29, 2014
Report
- Report Number
- 2939301-2014-18989
- Event Type
- Malfunction
- Date Received
- July 29, 2014
- Report Date
- July 21, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FOLLOW-UP # 1 ¿ (09/19/2014). THE LAY USER/PATIENT¿S METER HAS BEEN RETURNED ON 08/25/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 09/12/2014 WITH THE FOLLOWING FINDINGS:THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE SPC PINS 2 AND 3 LIFTED HIGH. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
Description of Event or Problem · 1
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR 5 METER ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 441193 | OT PING ENHANCED METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3612314 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |