COULTER LH 500 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2014-01749
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Date of Event
- July 3, 2014
- Report Date
- July 3, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K042724
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER TROUBLESHOOTED THE INSTRUMENT WITH BECKMAN COULTER (BEC) CUSTOMER TECHNICAL SUPPORT (CTS). THE CUSTOMER LOCATED CUT TUBING AT PV37; REPLACEMENT OF TUBING AT PV37 RESOLVED THE LEAK AND THE LYSE TEMPERATURE ERRORS. BEC FIELD SERVICE WAS NOT DISPATCHED FOR THIS EVENT. (B)(4).
THE CUSTOMER REPORTED A COULTER LH 500 HEMATOLOGY ANALYZER LEAKED BLUE FLUID DURING INSTRUMENT START UP. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. IN ADDITION, THE INSTRUMENT GENERATED LYSE TEMPERATURE ERRORS. THE INSTRUMENT OPERATOR WAS WEARING GLOVES, A FACE SHIELD, AND A LAB COAT WHEN THE LEAK WAS FOUND. THERE WERE NO REPORTS OF BIOHAZARD EXPOSURE TO THE LEAK. THERE WERE NO ERRONEOUS RESULTS GENERATED IN CONNECTION WITH THIS EVENT AND THERE WAS NO IMPACT TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 437847 | COULTER LH 500 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |