DIMENSION VISTA 1500
Report
- Report Number
- 1226181-2014-00404
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Date of Event
- July 3, 2014
- Report Date
- July 3, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC
- Product Code
- JGS
- PMA / PMN Number
- K051087
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CUSTOMER CONTACTED THE SIEMENS TECHNICAL SOLUTIONS CENTER (TSC). THE CUSTOMER STATED THAT THE SAMPLE RESULTED AS EXPECTED UPON REPEAT TESTING ON THE SAME INSTRUMENT. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED NA AND CL RESULTS IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT, FALSELY ELEVATED SODIUM (NA) AND CHLORIDE (CL) RESULTS WERE OBTAINED ON ONE PATIENT SAMPLE ON A DIMENSION VISTA 1500 INSTRUMENT. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED THREE TIMES ON THE SAME INSTRUMENT, RESULTING LOWER. IT IS UNKNOWN IF THE CORRECTED RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED NA AND CL RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 437650 | DIMENSION VISTA 1500 | CLINICAL CHEMISTRY ANALYZER | JGS | SIEMENS HEALTHCARE DIAGNOSTICS INC | DIMENSION VISTA 1500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |