ANIMAS VIBE
Report
- Report Number
- 2531779-2014-21617
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Report Date
- July 16, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1: DATE OF SUBMISSION 08/27/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE TEXT ON THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THE CONTRAST SETTING WAS INCREASED TO THE MAXIMUM WITH LITTLE IMPROVEMENT OBSERVED. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED. THERE WAS NO ISSUE WITH LOSS OF POWER AND NO EVIDENCE OF MOISTURE DAMAGE IN THE BATTERY COMPARTMENT.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 438118 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |