FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3961743 · Received July 28, 2014

Report

Report Number
2124215-2014-14530
Event Type
Injury
Date Received
July 28, 2014
Date of Event
June 6, 2014
Report Date
June 16, 2014
Manufacturer
CPI - DEL CARIBE
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, AN EVALUATION OF THE LEAD WAS PERFORMED. A VISUAL INSPECTION OF THE LEAD PROXIMAL SEGMENT RETURNED INDICATES THAT THE LEAD WAS SEVERED FROM THE TERMINAL PIN. THE OBSERVATION AGAINST THE LEAD COULD NOT BE CONFIRMED BY ANALYSIS. THE LEAD SEGMENT PASSED ELECTRICAL TESTING. FURTHER, THE DAMAGE TO THE LEAD WAS CONSISTENT WITH DAMAGE CAUSED DURING EXPLANT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WITH THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW OUT-OF-RANGE (OOR) PACING LEAD IMPEDANCE. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED TROUBLESHOOTING OPTIONS AND RECOMMENDED BRINGING THE PATIENT IN FOR EVALUATION. ADDITIONAL INFORMATION INDICATES THAT THERE WAS NOISE ON THE LEAD AND A LEAD FRACTURE WAS SUSPECTED HOWEVER, WAS NOT CONCLUSIVELY DETERMINED. THE RV WAS SURGICALLY ABANDONED AND THE DEVICE WAS EXPLANTED AND NO LONGER IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
439238 ENDOTAK RELIANCE IMPLANTABLE LEAD LWS CPI - DEL CARIBE 0184

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| L| R E110| 438-05| A155| 438-07| 4244| 0145| 294-09| 1831| 0184