FDA Adverse Event Malfunction Summary report: N

75MM SELECTABLE NEW TLC

MDR report key: 3961736 · Received July 28, 2014

Report

Report Number
3005075853-2014-05265
Event Type
Malfunction
Date Received
July 28, 2014
Date of Event
June 6, 2014
Report Date
June 11, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K092577
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE NTLC75 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AS THE FIRING KNOB WAS DETACHED AND THE SLIP BLOCK ASSEMBLY WAS NOTED TO BE DAMAGED. THE DEVICE WAS RECEIVED WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RETURNED FULLY FIRED. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TEST COULD BE PERFORMED. THE DAMAGE TO THE FIRING KNOB AND SLIP BLOCK ASSEMBLY IS CONSISTENT WITH HIGH (OUTSIDE INDICATED USE) STAPLE FORMING FORCES; HOWEVER THERE IS INSUFFICIENT EVIDENCE TO DETERMINE THE CAUSE OF THE HIGHER LOADS. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A RIGHT COLECTOMY PROCEDURE, THE DEVICE WAS FIRED FOR FIFTH TIME AND AS THE FIRING KNOB WAS RETURNED TO IT'S ORIGINAL POSITION, THE FIRING KNOB BROKE OFF AT THE CURVE OF HANDLE. THE FIRING KNOB DID NOT FALL INTO PATIENT. THE CASE WAS COMPLETED AT THAT POINT, SO NO NEW DEVICE WAS NEEDED. THERE WERE NO PATIENT CONSEQUENCES REPORTED.