FDA Adverse Event Injury Summary report: N

PROLITE MESH

MDR report key: 3961721 · Received June 16, 2014

Report

Report Number
1219977-2014-00207
Event Type
Injury
Date Received
June 16, 2014
Report Date
July 1, 2014
Manufacturer
ATRIUM MEDICAL CORP.
Product Code
FTL
PMA / PMN Number
K930669
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE MANUFACTURING LOT HISTORY AND STERILIZATION RECORDS WAS CONDUCTED. ALL IN-PROCESS SPECIFICATIONS AND RELEASE CRITERIA WERE MET. BALL BURST AND SUTURE RETENTION (COURSE AND WALE DIRECTIONS) WERE CONDUCTED ON THE LOTS OF POLYPROPYLENE MESH MATERIAL AT INCOMING AND ALL ACCEPTANCE CRITERIA WAS ALSO MET. POUCH SEAL STRENGTH TESTING WAS PERFORMED IN-PROCESS PRIOR TO RELEASE OF PRODUCT TO THE STERILIZER AND ALL DATA PASSED PER QC REVIEW.

Additional Manufacturer Narrative · 1

A THOROUGH INVESTIGATION WAS NOT ABLE TO BE PERFORMED AS NO PRODUCT CODE, LOT NUMBER OR SAMPLE WAS PROVIDED. A REVIEW OF COMPLAINTS WAS PERFORMED AND THERE HAVE NOT BEEN ANY SIMILAR REPORTS RELATED TO A DEVICE MALFUNCTION. THIS REPORT IS BASED UPON ALLEGATIONS MADE IN A LAWSUIT IN WHICH ATRIUM MEDICAL IS NAMED DEFENDANT. THIS REPORT SHALL NOT BE CONSIDERED AS AN ADMISSION BY ATRIUM MEDICAL THAT THE PRODUCT DESCRIBED IN THE LAWSUIT AND DESCRIBED HEREIN ARE OR WERE DEFECTIVE, OR THAT IT HAD ANY CAUSAL RELATIONSHIP TO ANY INJURIES ALLEGEDLY SUFFERED BY THE PLAINTIFF.

Description of Event or Problem · 1

THIS EVENT IS SEEMED REPORTABLE BASED ON THE ALLEGATIONS IN A LAWSUIT WHICH, WHILE UNSUBSTANTIATED, SUGGEST THAT A REPORTABLE EVENT MAY HAVE OCCURRED DURING USE OF ATRIUM MEDICALS MESH PRODUCT. PLAINTIFF ALLEGES MESH IMPLANTED AND INJURIES. SINCE THIS IS A LEGAL MATTER, THE CASE HAS BEEN TURNED OVER TO LEGAL COUNSEL AND FURTHER INFO OBTAINED THROUGH INVESTIGATION OR DISCOVERY MAY FALL UNDER THE ATTORNEY/CLIENT AND/OR WORK PRODUCT PRIVILEGE. HOWEVER, ATRIUM WILL SUPPLEMENT THIS REPORT AS APPROPRIATE IF ADDITIONAL INFO IS OBTAINED A FOLLOW UP REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353272 PROLITE MESH MESH, SURGICAL, POLYMERIC FTL ATRIUM MEDICAL CORP. 1010306-02 10426937

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention