FDA Adverse Event Injury Summary report: N

3DKNEE

MDR report key: 3961716 · Received July 28, 2014

Report

Report Number
1644408-2014-00483
Event Type
Injury
Date Received
July 28, 2014
Date of Event
July 25, 2014
Report Date
July 25, 2014
Manufacturer
ENCORE MEDICAL, L.P.
Product Code
JWH
PMA / PMN Number
K020114
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REASON FOR THIS REVISION WAS INSTABILITY OF THE JOINT AFTER 5.6 YEARS OF PATIENT USE. NO INFORMATION WAS REPORTED WITH THIS COMPLAINT ABOUT ANY OTHER PATIENT INJURIES, ACTIVITIES, ACCIDENTS, OR MEDICAL CONTRAINDICATIONS THAT MAY HAVE CONTRIBUTED TO THE NEED FOR THIS REVISION SURGERY. THE HEALTHCARE PROFESSIONAL INDICATED A SIGNIFICANT ADVERSE EVENT TO THE PATIENT. THERE WAS NO DELAY IN SURGERY AND ANOTHER SUITABLE DEVICE WAS AVAILABLE FOR USE. THE REVISION SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS DISPOSED OF AT THE HOSPITAL AND NOT MADE AVAILABLE TO DJO SURGICAL FOR EXAMINATION. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED NO NON-CONFORMING MATERIAL REPORTS ASSOCIATED WITH THIS PRODUCT. A REVIEW OF THE PRODUCT COMPLAINT REPORT HISTORY SHOWED THIS IS THE (B)(4) COMPLAINT AGAINST LOT NUMBER 53813313. A TOTAL OF (B)(4) COMPLAINTS HAVE BEEN SUBMITTED AGAINST THIS PART NUMBER. (B)(4) OF THE PREVIOUS INVESTIGATIONS HAVE BEEN ATTRIBUTABLE TO PATIENT JOINT INSTABILITY WITH THE REMAINDER RESULTING FROM ISSUES NOT ASSOCIATED TO LIKE EVENTS. THE ROOT CAUSE OF THIS EVENT IS SOFT TISSUE IN THE JOINT LEADING TO JOINT INSTABILITY. THERE IS NO REPORTED INFORMATION THAT INDICATES A MATERIAL, DESIGN, OR MANUFACTURING PROBLEM WITH THIS PRODUCT.

Description of Event or Problem · 1

REVISION SURGERY - DUE TO SOFT TISSUE AND INSTABILITY. THE SURGEON DECIDED TO USE A THICKER INSERT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
439196 3DKNEE 3D DCM INSERT, 9MM 4 RIGHT JWH ENCORE MEDICAL, L.P. 53813313

Patients

Seq Age Sex Outcome Treatment
1 68 YR Other| R