ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-05252
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Date of Event
- July 14, 2014
- Report Date
- July 15, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND IT WAS RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED, EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN11 AN ALERT SCREEN WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN ALERT SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE.
IT WAS REPORTED THAT THE TITANIUM TIP BROKE DURING AN UNKNOWN PROCEDURE. THE BROKEN PIECE WAS RETRIEVED BY FORCEPS. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 439710 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |