FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3961556 · Received July 28, 2014

Report

Report Number
1416980-2014-24471
Event Type
Malfunction
Date Received
July 28, 2014
Date of Event
May 22, 2014
Report Date
July 3, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS EVALUATED AND THE REPORTED ISSUE WAS VERIFIED THROUGH THE REVIEW OF THE EVENT HISTORY LOGS. THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 11:36:51. DURING NIGHT DRAIN CYCLE ONE, THE PATIENT'S ULTRAFILTRATION READING WAS 1791ML, INDICATING THE HOME PATIENT (HP) DRAINED 1791ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2400ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
437943 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 89 YR