PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88248
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 13, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
PUMP UNABLE TO PRIME DURING THE PRIME/A33 TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. NO A33 ALARM NOTED. UNIT HAD SCRATCHED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNER, AND CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED THAT CUSTOMER WAS HAVING ISSUES WITH HIS INSULIN PUMP. HIS BLOOD GLUCOSE IS 9.7 MMOL/L. CUSTOMER'S DEVICE IS ALARMING. HE IS UNABLE TO EXIT THE PREPARING TO PRIME LOOP. CUSTOMER ALSO SAID HE LOST HIS BELT CLIP, SO THE DEVICE WAS DROPPED ONCE PRIOR TO THE ALARM. THE DRIVE SUPPORT CAP APPEARED NORMAL. CUSTOMER WAS ADVISED THAT THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. HE WAS ADVISED THE POSSIBLE CAUSE OF THE ALARM WAS THAT DURING SEATING THE FORCE SENSOR DID NOT DETECT THE RESERVOIR. NOTHING FURTHER REPORTED.