FDA Adverse Event Injury Summary report: N

TALENT

MDR report key: 3961389 · Received July 28, 2014

Report

Report Number
2953200-2014-01469
Event Type
Injury
Date Received
July 28, 2014
Report Date
July 11, 2014
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P070027
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE TALENT BIFURCATE HAS MIGRATED APPROXIMATELY 4-5CM AND IS IN THE ANEURYSM SAC. A PROXIMAL TYPE I ENDOLEAK WAS ALSO OBSERVED. THE PHYSICIAN SUSPECTS THE CAUSE OF THE EVENT WAS MOST LIKELY DUE TO DISEASE PROGRESSION; NECK DILATATION. THE PHYSICIAN INTERVENED AND REALIGNED THE TALENT STENT GRAFT WITH AN ENDURANT BIFURCATE AND CONTRALATERAL LIMBS RESOLVING THE ENDOLEAK. THERE WERE NO ISSUES DURING THE INTERVENTION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
437387 TALENT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA V00176006

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Required Intervention