FDA Adverse Event Malfunction Summary report: N

CARE ASSIST BED

MDR report key: 3961158 · Received May 28, 2014

Report

Report Number
9615739-2014-00223
Event Type
Malfunction
Date Received
May 28, 2014
Date of Event
May 2, 2014
Report Date
May 2, 2014
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BRAKE CASTERS NOT HOLDING DUE TO NORMAL WEAR AND TEAR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012 - 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECH REPLACED THE TWO BRAKES AND ONE STEER CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECH STATING THE BRAKE CASTERS WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN THE BED SHOP. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314930 CARE ASSIST BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1