HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-24435
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 7, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL AND MICROSCOPIC INSPECTION REVEALED BURNT PLASTIC PARTICULATE MATTER EMBEDDED INSIDE THE TUBING WALL OF THE PATIENT LINE. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A MANUFACTURING ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THE SAMPLE HAS BEEN RECEIVED BUT THE EVALUATION HAS YET TO BE COMPLETED. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS CONDUCTED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT BLACK SPOTS WERE OBSERVED ON THE INSIDE OF THE TUBING OF THE CASSETTE. THE EVENT WAS REPORTED BEFORE USE ON THE HOME CHOICE (HC). NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 439635 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE | S14B28084 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |