FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3961097 · Received July 15, 2014

Report

Report Number
1627487-2014-12472
Event Type
Injury
Date Received
July 15, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WANT OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 3 OF 3. REFER TO MFR REPORT #1627487-2014-12471, 1627487-2014-12470. IT WAS REPORTED THAT THE PATIENT'S IPG IS TENDER TO THE TOUCH AND THE IPG IS MOVING WHEN SHE WALKS. THE PATIENT EXPERIENCED SHOCKING AT RANDOM TIMES. THE PATIENT REPORTED THAT THE STIMULATION TURNS ON AND OFF BY ITSELF. THE SCS SYSTEM REQUIRES RECHARGING MORE FREQUENTLY AND EXPERIENCES HEATING WHILE CHARGING. SURGICAL INTERVENTION IS PLANNED TO ADDRESS THE ISSUE. ON 08/01/2012 ST JUDE MEDICAL NEUROMODULATION DIVISION SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413230 CHARGING SYSTEM SCS CHARGER GZB ST JUDE MEDICAL - NEUROMODULATION 3721 3233919

Patients

Seq Age Sex Outcome Treatment
1 51 YR Other