THUNDERBEAT 5MM 35CM FRONT-ACTUATED GRIP
Report
- Report Number
- 8010047-2014-00279
- Event Type
- Malfunction
- Date Received
- May 26, 2014
- Date of Event
- May 9, 2014
- Report Date
- May 9, 2014
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PTFE PAD PARTIALLY WORN AWAY. THERE WERE CONTACT MARKS ON THE SURFACE OF THE PROBE AND THE JAW. THE MANUFACTURING RECORD WAS REVIEWED WITH NO IRREGULARITIES. BASED ON THE SIMILAR CASES WITH THE SAME MODEL, IT IS KNOWN THAT BY CONTINUOUSLY ACTIVATING OUTPUT WITHOUT GRASPING ANYTHING IN THE GRASPING SECTION (INCLUDING AFTER THE TISSUE SEPARATED), THE PTFE PAD SEVERELY WEARS. CONSIDERING THE EVALUATION RESULT OF THE SUBJECT DEVICE, OMSC CONCLUDED THAT THE REPORTED EVENT OCCURRED SINCE THE USER CONTINUED ACTIVATING OUTPUT FOR AN EXTENDED TIME AFTER THE TISSUE ALREADY CUT. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION. SEE SCANNED PAGE.
DURING LAPAROSCOPIC ASSISTED DISTAL GASTRECTOMY, THE PTFE PAD OF THE SUBJECT DEVICE WAS DEFORMED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT INJURY WITHOUT THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310054 | THUNDERBEAT 5MM 35CM FRONT-ACTUATED GRIP | THUNDERBEAT HANDPIECE | GEI | OLYMPUS MEDICAL SYSTEMS CORPORATION | TB-0535FC | K3925 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |