FDA Adverse Event Malfunction Summary report: N

THUNDERBEAT 5MM 35CM FRONT-ACTUATED GRIP

MDR report key: 3961081 · Received May 26, 2014

Report

Report Number
8010047-2014-00279
Event Type
Malfunction
Date Received
May 26, 2014
Date of Event
May 9, 2014
Report Date
May 9, 2014
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PTFE PAD PARTIALLY WORN AWAY. THERE WERE CONTACT MARKS ON THE SURFACE OF THE PROBE AND THE JAW. THE MANUFACTURING RECORD WAS REVIEWED WITH NO IRREGULARITIES. BASED ON THE SIMILAR CASES WITH THE SAME MODEL, IT IS KNOWN THAT BY CONTINUOUSLY ACTIVATING OUTPUT WITHOUT GRASPING ANYTHING IN THE GRASPING SECTION (INCLUDING AFTER THE TISSUE SEPARATED), THE PTFE PAD SEVERELY WEARS. CONSIDERING THE EVALUATION RESULT OF THE SUBJECT DEVICE, OMSC CONCLUDED THAT THE REPORTED EVENT OCCURRED SINCE THE USER CONTINUED ACTIVATING OUTPUT FOR AN EXTENDED TIME AFTER THE TISSUE ALREADY CUT. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION. SEE SCANNED PAGE.

Description of Event or Problem · 1

DURING LAPAROSCOPIC ASSISTED DISTAL GASTRECTOMY, THE PTFE PAD OF THE SUBJECT DEVICE WAS DEFORMED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT INJURY WITHOUT THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310054 THUNDERBEAT 5MM 35CM FRONT-ACTUATED GRIP THUNDERBEAT HANDPIECE GEI OLYMPUS MEDICAL SYSTEMS CORPORATION TB-0535FC K3925

Patients

Seq Age Sex Outcome Treatment
1