FDA Adverse Event Malfunction Summary report: N

ACCESS

MDR report key: 3960944 · Received July 28, 2014

Report

Report Number
1416980-2014-24427
Event Type
Malfunction
Date Received
July 28, 2014
Report Date
July 3, 2014
Manufacturer
BIEFFE MEDITAL MANUFACTURING S.A.R.L. TUNIS
Product Code
FPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE USED DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. GRAVITY TESTING WAS PERFORMED, AND THE LIQUID WAS NOTED TO FLOW CORRECTLY THROUGH THE SET WITH NO LEAKS DETECTED. THE DEVICE OPERATED PER SPECIFICATION. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A CHEMO-AIDE CONTINU-LINE SET LEAKED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
439460 ACCESS SET, ADMINISTRATION, INTRAVASCULAR FPA BIEFFE MEDITAL MANUFACTURING S.A.R.L. TUNIS 13G19V111