UNIPLANAR SCREW (TI-6AL-4V)
Report
- Report Number
- 2027467-2014-00021
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Date of Event
- June 5, 2014
- Report Date
- July 3, 2014
- Manufacturer
- ALPHATEC SPINE INC
- Product Code
- KWP
- PMA / PMN Number
- K100685
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER
Narratives
NO PRODUCT EVALUATION CAN BE CONDUCTED. THE SPLAYED/BENT UNIPLANAR SCREW REMAINS SECURED WITHIN THE PATIENTS L1 VERTEBRAL BODY. THE IDENTIFYING PART AND LOT NUMBERS HAVE NOT BEEN PROVIDED. BASED ON THE SUPPLIED PHOTO AND WHAT WAS DESCRIBED IN THE REPORT, IT APPEARS THE AXIAL REDUCER MAY NOT HAVE BEEN COMPLETELY ATTACHED AND WHEN THE FORCES WERE PLACED ON THE SCREW DURING THE DE-ROTATION MANEUVER. IF THE INSTRUMENT (AXIAL REDUCER) IS NOT ATTACHED APPROPRIATELY WITH GOOD ALIGNMENT THE SETSCREW MAY BE SLIGHTLY MISALIGNED WHICH COULD POTENTIALLY LEAD TO A CROSS THREADING OF THE SETSCREW. THE ZODIAC SPINAL FIXATION SYSTEM IS INTENDED FOR USE AS A POSTERIOR SPINAL FIXATION DEVICE TO AID IN THE SURGICAL CORRECTION OF VARIOUS SPINAL DEFORMITIES AND PATHOLOGIES OF THE SPINE. IT IS INTENDED TO PROVIDE STABILIZATION DURING THE DEVELOPMENT OF FUSION UTILIZING A BONE GRAFT. SPECIFIC INDICATIONS FOR THE ZODIAC SYSTEM ARE DEPENDENT IN PART ON THE CONFIGURATION OF THE ASSEMBLED DEVICE AND THE METHOD OF ATTACHMENT TO THE SPINE.
AN INTERNATIONAL CUSTOMER (B)(6) REPORTED AN INCIDENT INVOLVING A ZODIAC UNIPLANAR SCREW. THE EVENT OCCURRED DURING A DEFORMITY (IDIOPATHIC SCOLIOSIS) CASE ON (B)(6) 2014. DURING THE FINAL TIGHTENING PROCESS, THE HEAD/TULIP OF THE ZODIAC UNIPLANAR SCREW LOCATED AT THE L1 BENT AND CRACKED PREVENTING TORQUING OF THE SET SCREW. THE SPLAYED HEAD/TULIP DID NOT DETACH FROM THE BODY OF THE UNIPLANAR SCREW, IT REMAINS POSITIONED AT THE L1 ENTRAPPED BENEATH THE ROD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 437604 | UNIPLANAR SCREW (TI-6AL-4V) | KWP, MNI, MNH | KWP | ALPHATEC SPINE INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |