FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3960879 · Received June 27, 2014

Report

Report Number
1720753-2014-05539
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 11, 2014
Report Date
June 27, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FFB BOARD WAS EVALUATED AND REPLACED. A COMPLETE FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
375937 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1