OLYMPUS RESECTION SHEATH
Report
- Report Number
- 2951238-2014-00281
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 4, 2014
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- FJL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
THE DEVICE REFERENCE IN THIS REPORT WAS RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION CONFIRMED THAT A PORTION OF THE CERAMIC BEAK AT THE DISTAL TIP OF THE RESECTION SHEATH BROKE OFF. THE DEVICE WAS SERVICED AND RETURNED TO THE FACILITY. THE DAMAGE FOUND ON THE DEVICE IS CONSISTENT WITH PHYSICAL DAMAGE. THE DEVICE INSTRUCTION MANUAL STATES: "VISUALLY INSPECT THE CERAMIC INSULATION AT THE SHEATH'S DISTAL END BEFORE EACH USE. DO NOT USE THE INSTRUMENT IN CASE OF DAMAGE (E.G. CRACKS. FRACTURES)." ALSO, WARNING STATEMENT: "IMPACT, FALL, SHOCK OR SIMILAR STRESS CAN DAMAGE THE CERAMIC INSULATION."
OLYMPUS WAS INFORMED THAT TOWARDS THE END OF A TRANSURETHRAL RESECTION OF THE BLADDER (TURB), THE CERAMIC TIP OF THE RESECTION SHEATH BROKE OFF AND FELL INSIDE THE PT'S BLADDER. THE BROKEN TIP WAS RECOVERED FROM THE PT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 376159 | OLYMPUS RESECTION SHEATH | RESECTION SHEATH | FJL | OLYMPUS WINTER & IBE GMBH | A22041A | 138W-0015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |