FDA Adverse Event Malfunction Summary report: N

OLYMPUS RESECTION SHEATH

MDR report key: 3960851 · Received June 27, 2014

Report

Report Number
2951238-2014-00281
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 4, 2014
Report Date
June 4, 2014
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
FJL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE REFERENCE IN THIS REPORT WAS RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION CONFIRMED THAT A PORTION OF THE CERAMIC BEAK AT THE DISTAL TIP OF THE RESECTION SHEATH BROKE OFF. THE DEVICE WAS SERVICED AND RETURNED TO THE FACILITY. THE DAMAGE FOUND ON THE DEVICE IS CONSISTENT WITH PHYSICAL DAMAGE. THE DEVICE INSTRUCTION MANUAL STATES: "VISUALLY INSPECT THE CERAMIC INSULATION AT THE SHEATH'S DISTAL END BEFORE EACH USE. DO NOT USE THE INSTRUMENT IN CASE OF DAMAGE (E.G. CRACKS. FRACTURES)." ALSO, WARNING STATEMENT: "IMPACT, FALL, SHOCK OR SIMILAR STRESS CAN DAMAGE THE CERAMIC INSULATION."

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT TOWARDS THE END OF A TRANSURETHRAL RESECTION OF THE BLADDER (TURB), THE CERAMIC TIP OF THE RESECTION SHEATH BROKE OFF AND FELL INSIDE THE PT'S BLADDER. THE BROKEN TIP WAS RECOVERED FROM THE PT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376159 OLYMPUS RESECTION SHEATH RESECTION SHEATH FJL OLYMPUS WINTER & IBE GMBH A22041A 138W-0015

Patients

Seq Age Sex Outcome Treatment
1