FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 3960819 · Received June 27, 2014

Report

Report Number
1720753-2014-05547
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 12, 2014
Report Date
June 27, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE MONO-BLOCK X-RAY WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXPERIENCED AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
375863 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 8800

Patients

Seq Age Sex Outcome Treatment
1