FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3960732
·
Received June 9, 2014
Report
- Report Number
- 1720753-2014-04817
- Event Type
- Malfunction
- Date Received
- June 9, 2014
- Date of Event
- May 20, 2014
- Report Date
- June 9, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE WAS FORMATTED AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A CINE DISK NOT AVAILABLE ERROR MESSAGE. THIS WOULD PREVENT THE CINE FUNCTION FROM OPERATING, CAUSING A LOSS OF SUBTRACTION, ROAD-MAPPING, OR OTHER CRITICAL VASCULAR FUNCTIONS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333830 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |