ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2014-21566
- Event Type
- Malfunction
- Date Received
- July 28, 2014
- Report Date
- July 15, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. THE VIBRATION MOTOR WAS UNRESPONSIVE DURING TESTING. THE PUMP COVER WAS OPENED AND THE VIBRATION MOTOR PINS WERE MISALIGNED. UNRELATED TO THESE ISSUES, THE PUMP¿S PAINT WAS FOUND TO BE WORN. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DISPLAY FAILURE AND A VIBRATION MOTOR FAILURE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 437658 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |