FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3960719 · Received June 9, 2014

Report

Report Number
1720753-2014-04826
Event Type
Malfunction
Date Received
June 9, 2014
Date of Event
May 20, 2014
Report Date
June 9, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE BIOS WERE RESET. THE GENERAL PURPOSE OPERATING SYSTEM (GPOS) AND THE SOFTWARE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED INTERMITTENT COMMUNICATION FAILED ERROR MESSAGES. THIS ERROR RESULTS IN X-RAYS BEING DISABLED. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333600 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1