FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 3960584 · Received July 28, 2014

Report

Report Number
1031452-2014-04392
Event Type
Malfunction
Date Received
July 28, 2014
Date of Event
June 13, 2014
Report Date
June 20, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

UNIT ALARMING PER PROVIDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
439478 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other