FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3960441 · Received June 9, 2014

Report

Report Number
1720753-2014-04851
Event Type
Malfunction
Date Received
June 9, 2014
Date of Event
May 20, 2014
Report Date
June 9, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE HV CABLE CONNECTIONS X-RAY TUBE WERE EVALUATED, RE-LUBRICATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS AN 'SATURATION FAULT ERROR' MESSAGE DISPLAYED. THIS ERROR IS LIKELY RESULT IN AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335422 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1