FDA Adverse Event Malfunction Summary report: N

PLUM A+ 3 MEDNET SW

MDR report key: 3960367 · Received June 23, 2014

Report

Report Number
9615050-2014-04132
Event Type
Malfunction
Date Received
June 23, 2014
Date of Event
January 1, 2014
Report Date
May 20, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
FA210-07
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, CHANNEL 3 OF THE DEVICE ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CTR FOR RECERTIFICATION. DURING VERIFICATION TESTING AT THE SERVICE CTR, CHANNEL 3 OF THE DEVICE ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE. THIS ERROR OCCURS WHEN THE BATTERY CANNOT BE FULLY CHARGED WITHIN EIGHT HOURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366259 PLUM A+ 3 MEDNET SW 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA