FDA Adverse Event Malfunction Summary report: N

AVIATOR ASSY ONE LEVEL PLATE SIZE 12

MDR report key: 3960216 · Received July 28, 2014

Report

Report Number
0009617544-2014-00331
Event Type
Malfunction
Date Received
July 28, 2014
Date of Event
June 30, 2014
Report Date
June 30, 2014
Manufacturer
STRYKER SPINE-FRANCE
Product Code
KWQ
PMA / PMN Number
K083562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: DEVICE HISTORY REVIEW; RESULTS: MANUFACTURING FILES COULD NOT BE REVIEWED BECAUSE NO LOT NUMBER WAS PROVIDED AND NO PART WAS RETURNED FOR VISUAL EXAMINATION. THERE WAS NO DELAY IN SURGERY AND THE PLATE WAS SUCCESSFULLY INSTALLED USING A 14 MM HAND DRILL GUIDE. CONCLUSION: NO SPECIFIC DETAILS WERE PROVIDED CONCERNING THE TECHNIQUE USED, SUCH AS OVER-ARTICULATION OF THE DRILL GUIDE DURING INSTALLATION OR REMOVAL OF THE DRILL GUIDE. THEREFORE, THE EXACT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE IS MULTIFACTORIAL IN NATURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A C3-4 ACDF PROCEDURE WAS PERFORMED. A 12MM AVIATOR PLATE WAS USED AND THE SURGEON HAND DRILLED USING THE 14MM DRILL/GUIDE. WHEN INSERTING THE RIGHT C3 SCREW (4.0 X 14 VST), THE SCREW HEAD BECAUSE CAUGHT THE LOCKING FLANGE. THE FLANGE WAS BURIED BELOW THE SCREW. SURGEON DECIDED TO LEAVE SCREW IN AS THE PURCHASE WAS VERY SOLID AND NOTHING WAS PROMINENT. ALL 3 SCREWS WERE LOCKED WITH SECONDARY LOCKING MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT A C3-4 ACDF PROCEDURE WAS PERFORMED. A 12MM AVIATOR PLATE WAS USED AND THE SURGEON HAND DRILLED USING THE 14MM DRILL/GUIDE. WHEN INSERTING THE RIGHT C3 SCREW (4.0 X 14 VST), THE SCREW HEAD BECAUSE CAUGHT THE LOCKING FLANGE. THE FLANGE WAS BURIED BELOW THE SCREW. SURGEON DECIDED TO LEAVE SCREW IN AS THE PURCHASE WAS VERY SOLID AND NOTHING WAS PROMINENT. ALL 3 SCREWS WERE LOCKED WITH SECONDARY LOCKING MECHANISM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
437792 AVIATOR ASSY ONE LEVEL PLATE SIZE 12 IMPLANT-AVIATOR PLATE KWQ STRYKER SPINE-FRANCE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 59 YR