FDA Adverse Event Other Summary report: N

FUJINON

MDR report key: 3958711 · Received June 20, 2014

Report

Report Number
2431293-2014-00002
Event Type
Other
Date Received
June 20, 2014
Date of Event
January 1, 2014
Report Date
May 20, 2014
Manufacturer
FUJIFILM OPTICAL CORPORATION, MITO FACTORY
Product Code
FDT
PMA / PMN Number
K042076
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FUJIFILM MEDICAL SYSTEMS U.S.A., INC. (FMSU) IMMEDIATELY INITIATED AN INVESTIGATION, INCLUDING A VISIT TO (B)(6), TO IDENTIFY THE ROOT CAUSE FOR THE INCIDENT. FMSU PERSONNEL WERE INFORMED THAT, IN RESPONSE TO CRE INCIDENTS AT ITS SISTER HOSPITAL, (B)(6) HAD CONDUCTED A REVIEW OF PRIOR MEDICAL RECORDS FOR ITS PATIENT WHO WERE POSITIVE FOR CRE. AT THE TIME OF THIS REPORT, FMSU HAS BEEN ADVISED THAT TWO OF THREE ENDOSCOPES ((B)(4)) HAVE ALLEGEDLY BEEN CULTURED POSITIVE FOR CRE. THREE PATIENT WHO HAD UNDERGONE ERCP PROCEDURES TESTED POSITIVE FOR CRE AND TWO OF THREE OF THESE PATIENTS EXPRESSED SYMPTOMS CONSISTENT WITH POSITIVE CRE. THE CUSTOMER IS UNSURE WHETHER THE PATIENT TRANSFERRED CRE TO THE ENDOSCOPE OR VICE VERSA. NON-FUJIFILM EQUIPMENT, INCLUDING A CHANNEL CLEANING DEVICE INSTEAD OF CLEANING BRUSH, IS USED TO MANUALLY CLEAN THE ENDOSCOPES AND AN AER IS USED FOR AUTOMATED HIGH-LEVEL DISINFECTION. A REVIEW OF FMSU SERVICE RECORDS FOR (B)(6) INDICATED NO ABNORMALITIES OTHER THAN GENERAL WEAR AND TEAR REPAIRS ATTRIBUTABLE TO NORMAL USAGE AND HANDLING OF THE DEVICE. FMSU HAS REQUESTED BUT HAS NOT RECEIVED ANY INFORMATION ON ANY TREATMENT OR HOSPITALIZATIONS FOR THESE PATIENTS, AT THE TIME OF THIS INITIAL REPORT. IN ADDITION, FMSU HAS REQUESTED THAT THE DUODENOSCOPES THAT TESTED POSITIVE FOR CRE BE RETURNED FOR A DETAILED EXAMINATION. TO DATE, THE CUSTOMER HAS NOT RETURNED ANY DUODENOSCOPES TO FMSU. THE INVESTIGATION IS STILL ONGOING. FMSU WILL SUBMIT A SUPPLEMENTAL MDR ONCE THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

ON (B)(6) 2014, FUJIFILM MEDICAL SYSTEMS U.S.A., INC. (FMSU) WAS CONTACTED BY (B)(6) HOSPITAL AND ADVISED ON FUJINON (NOTE: FMSU'S ENDOSCOPES BEAR THE BRAND NAME OF "FUJINON") DUODENOSCOPES AND PATIENTS TESTING POSITIVE FOR CRE (CARBAPENEM-RESISTANT ENTEROBACTERIACEAE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
363374 FUJINON DUODENOSCOPE FDT FUJIFILM OPTICAL CORPORATION, MITO FACTORY ED-530XT

Patients

Seq Age Sex Outcome Treatment
1 Other