ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2014-21325
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Report Date
- July 21, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1: DATE OF SUBMISSION 10/16/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/02/2014 WITH THE FOLLOWING FINDINGS:ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE PUMP POWERED UP TO THE VERIFY SCREEN WITH AUDITORY AND VIBRATORY FEATURES. NO TACTILE ISSUES WERE FOUND WITH THE KEYPAD BUTTONS. REMOVAL OF THE KEYPAD COVER DID NOT FIND CONTAMINATION ON ANY OF THE BUTTON CONTACTS. THE REPORTED KEYPAD BUTTON TACTILE ISSUE WAS NOT DUPLICATED IN THE INVESTIGATION. UNRELATED TO THE REPORTED BUTTON ISSUES, THE DISPLAY WAS FOUND TO BE DIM AND DISCOLORED AND THE BOLUS BUTTON COVER WAS FOUND DETACHED/MISSING.
ON (B)(6) 2014, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING THAT THE ¿UP¿ AND ¿DOWN¿ ARROW BUTTONS WERE UNDER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434935 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |