FDA Adverse Event Malfunction Summary report: N

SMARTTORQUE S609 C

MDR report key: 3955791 · Received June 25, 2014

Report

Report Number
3003637274-2014-00023
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
May 22, 2014
Report Date
June 5, 2014
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS DID SHOW THAT THE BEARING HAVE BEEN GRITTY. THE BACK CAP HAD A DENT WHICH CAUSED THE BACK CAP BUTTON TO STICK IN. IF THE BACK CAP BUTTON STICKS IN IT CAUSES UNUSUAL INNER FRICTION AND THEREFORE A FAST INCREASE OF TEMPERATURE. THE DENT SHOWS THAT THE HANDPIECE RECEIVED A STRONG HIT, E.G. DROPPING DURING REPROCESSING. THE CONDITION OF THE BEARINGS IS DUE TO WEAR PROCESS. IT IS LIKELY THAT DUE TO THE HIGHER TEMPERATURE A STRONGER WEAR PROCESS TOOK PLACE. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.

Description of Event or Problem · 1

DURING A STANDARD DENTAL TREATMENT ON TOOTH #4 THE HANDPIECE HEATED UP AND CAUSED A BURN ON CHEEK TO ABOUT THE SIZE OF THE BACK CAP. THE PT WAS INSTRUCTED TO DO WARM SALT/WATER RINSES BUT DID NOT GET ANY MEDICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
370613 SMARTTORQUE S609 C DENTAL HANDPIECE EFB KAVO DENTAL GMBH S609 C

Patients

Seq Age Sex Outcome Treatment
1 21 YR