FDA Adverse Event
Malfunction
Summary report: N
SMARTTORQUE S609 C
MDR report key: 3955791
·
Received June 25, 2014
Report
- Report Number
- 3003637274-2014-00023
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- May 22, 2014
- Report Date
- June 5, 2014
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE ANALYSIS DID SHOW THAT THE BEARING HAVE BEEN GRITTY. THE BACK CAP HAD A DENT WHICH CAUSED THE BACK CAP BUTTON TO STICK IN. IF THE BACK CAP BUTTON STICKS IN IT CAUSES UNUSUAL INNER FRICTION AND THEREFORE A FAST INCREASE OF TEMPERATURE. THE DENT SHOWS THAT THE HANDPIECE RECEIVED A STRONG HIT, E.G. DROPPING DURING REPROCESSING. THE CONDITION OF THE BEARINGS IS DUE TO WEAR PROCESS. IT IS LIKELY THAT DUE TO THE HIGHER TEMPERATURE A STRONGER WEAR PROCESS TOOK PLACE. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.
Description of Event or Problem · 1
DURING A STANDARD DENTAL TREATMENT ON TOOTH #4 THE HANDPIECE HEATED UP AND CAUSED A BURN ON CHEEK TO ABOUT THE SIZE OF THE BACK CAP. THE PT WAS INSTRUCTED TO DO WARM SALT/WATER RINSES BUT DID NOT GET ANY MEDICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370613 | SMARTTORQUE S609 C | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | S609 C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |