FDA Adverse Event Injury Summary report: N

TRAILBLAZER SUPPORT CATHETER

MDR report key: 3955781 · Received July 24, 2014

Report

Report Number
2183870-2014-00188
Event Type
Injury
Date Received
July 24, 2014
Date of Event
June 27, 2014
Report Date
June 27, 2014
Manufacturer
COVIDIEN
Product Code
DQY
PMA / PMN Number
K092299
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW / REMEDIATION EFFORT PERFORMED AT THE COVIDIEN (B)(4) LOCATION, FOLLOWING MEDTRONIC¿S ACQUISITION OF COVIDIEN. A CAPA HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING.

Additional Manufacturer Narrative · 1

A REVIEW OF THE MANUFACTURE RECORDS FOR THIS DEVICE DID NOT REVEAL ANY DISCREPANCIES RELEVANT TO THE REPORTED EVENT. TWO ANGIOGRAPHIC IMAGES WERE RECEIVED SHOWING THE TRAILBLAZER IN THE VESSEL. ADDITIONAL INFORMATION HAS BEEN REQUESTED. SHOULD IT BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE TRAILBLAZER CATHETER WAS RECEIVED FOR EVALUATION WITHOUT THE SECTION OF THE CATHETER CONTAINING THE MARKER BANDS . THE DISTAL END OF THE FRACTURED CATHETER EXHIBITED LONGITUDINAL STRETCHING/ NECKING. THE LENGTH OF THE CATHETER SEGMENT WAS 132.8CM FROM THE DISTAL EDGE OF THE STRAIN RELIEF. THE NOMINAL DIMENSION FOR THE CATHETER TO THE PROXIMAL MARKER BAND IS APPROXIMATELY 105CM. THIS WOULD SUGGEST THAT THE CATHETER WAS SUBJECTED TO TENSILE FORCES EXCEEDING THE MATERIALS YIELD STRENGTH. THERE WAS ALSO A KINK 5.1CM FROM THE FRACTURE FACE.

Description of Event or Problem · 1

THIS PROCEDURE WAS PERFORMED IN (B)(6). WHILE WITHDRAWING THE TRAILBLAZER FROM THE PATIENT, THE PHYSICIAN FELT GREAT RESISTANCE WHEN IT PASSED THROUGH THE LESION POSITION. LOOKING AT ANGIO IMAGES THE TIP OF THE TRAILBLAZER WAS FOUND TO BE BROKEN AND LEFT IN THE PATIENT. THE BROKEN TIP OF THE TRAILBLAZER WAS JAILED UP WITH A STENT IN THE VESSEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434245 TRAILBLAZER SUPPORT CATHETER CATHETER, PERCUTANEOUS DQY COVIDIEN SC-018-135 9846810

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention