FDA Adverse Event
Malfunction
Summary report: N
VERSACARE BED
MDR report key: 3955727
·
Received February 18, 2014
Report
- Report Number
- 1824206-2014-00466
- Event Type
- Malfunction
- Date Received
- February 18, 2014
- Date of Event
- January 20, 2014
- Report Date
- January 20, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT'S CLINICAL ENGINEER FOUND THE BRAKE CASTER STEM WAS SNAPPED. A SEARCH FOR THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECH REPLACED THE BRAKE CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYS AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 101293 | VERSACARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |