ARTICULEZE M HEAD 36MM +5
Report
- Report Number
- 1818910-2014-24176
- Event Type
- Injury
- Date Received
- July 24, 2014
- Date of Event
- July 3, 2014
- Report Date
- July 3, 2014
- Manufacturer
- DEPUY INTL., LTD.8010379
- Product Code
- JDI
- PMA / PMN Number
- PK980513
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
UDI: UNAVAILABLE. DEPUY SYNTHES HAS BEEN INFORMED THAT THE CATALOG NUMBER AND LOT NUMBER IS NOT AVAILABLE. NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. NO FURTHER INFORMATION OR INVESTIGATIONAL INPUTS HAVE BEEN MADE AVAILABLE. A SEARCH OF THE COMPLAINTS DATABASES FINDS OTHER REPORTS AGAINST THE FEMORAL HEAD PRODUCT/LOT CODE COMBINATION. REVIEW OF DEVICE HISTORY RECORDS PREVIOUSLY FINDS NO RELATED MANUFACTURING DEVIATIONS OR ANOMALIES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE METAL LINER PRODUCT AND LOT CODE COMBINATION SINCE ITS RELEASE TO DISTRIBUTION. THE INVESTIGATION CAN DRAW NO CONCLUSION WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEPUY CONSIDERS THE INVESTIGATION COMPLETED. SHOULD FURTHER INFORMATION BE RECEIVED, THE COMPLAINT WILL BE UPDATED AT THAT TIME.
UPDATE REC'D 09/15/2015 - THE PATIENT'S MEDICAL RECORDS WERE RECEIVED. MEDICAL RECORDS WERE REVIEWED FOR MDR REPORTABILITY. ACCORDING TO THE MEDICAL RECORDS, THE PATIENT WAS EXPERIENCING PAIN AND UNDERWENT A CLOSED REDUCTION TO ADDRESS A DISLOCATION. UPON REVISION AN OSTEOLYTIC FRACTURE, PSEUDOTUMOR, AND CORROSION AT THE TRUNNION. AT THIS TIME THE PATIENT'S FEMORAL STEM IS BEING ADDED TO THE COMPLAINT AND REPORTED.
PATIENT WAS REVISED TO ADDRESS METALLOSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434292 | ARTICULEZE M HEAD 36MM +5 | HIP FEMORAL HEAD | JDI | DEPUY INTL., LTD.8010379 | 2163931 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention |