FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3955644
·
Received June 11, 2014
Report
- Report Number
- 8020893-2014-01411
- Event Type
- Malfunction
- Date Received
- June 11, 2014
- Date of Event
- January 1, 2014
- Report Date
- May 15, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER REPORTED TO HAVE REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB). THE COVIDIEN CUSTOMER SVC ENGINEER (CSE) UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. COVIDIEN REF: (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED BY A CUSTOMER IN THE USA THAT AN 840 VENTILATOR EXHIBITED AN ERRATIC BOTTOM DISPLAY. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 345079 | 840 VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |