FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

MDR report key: 3955632 · Received June 13, 2014

Report

Report Number
1419937-2014-00488
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
February 13, 2014
Report Date
February 13, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT TRANSFORMER WAS SENT TO THE CUSTOMER. THE PRODUCT INVOLVED IN THIS COMPLAINT WAS RETURNED FOR EVALUATION AND WAS DETERMINED TO BE BREACH. THE CAUSE OF THE BREACH IN THE REV N TRANSFORMER HAS NOT BEEN DETERMINED AT THIS TIME. IT IS CURRENTLY BEING INVESTIGATION UNDER (B)(4). EVALUATION SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS CRACKED OPEN. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS NOT ABLE TO BE PLUGGED IN SO THE VOLTAGE ACROSS THE DC PLUG WAS NOT MEASURED. RESISTANCE ACROSS THE DC PLUG WAS MEASURED .753M, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 56.7, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT HER TRANSFORMER HOUSING IS CRACKED. WHEN THE PRODUCT WAS RECEIVED AT MEDELA IT WAS DETERMINED TO BE BREACH, WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350772 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC. 57063/9207010 REV N-1413

Patients

Seq Age Sex Outcome Treatment
1