ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-21292
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Report Date
- July 18, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 09/02/2014 WITH THE FOLLOWING FINDINGS: THE INITIAL COMPLAINT COULD NOT BE INVESTIGATED DUE TO A DISPLAY FAILURE, THUS NO CONCLUSIONS CAN BE MADE IN THIS REGARD. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED AT THE THREADS AND CASE SEAL. THE PUMP POWERED ON, AS CONFIRMED BY THE PRESENCE OF THE AUDITORY AND VIBRATORY CUES, BUT THE DISPLAY SCREEN WAS BLANK. THE PUMP CASE WAS REMOVED, AND EVIDENCE OF CORROSION WAS OBSERVED ON THE DISPLAY AND ALONG THE BATTERY CYLINDER. THE SUSPECT DISPLAY SCREEN WAS REPLACED WITH A TEST DISPLAY SCREEN; THE PUMP DISPLAY FUNCTIONED PROPERLY WITH THE TEST DISPLAY SCREEN INSTALLED.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 435048 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |