FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 3955474 · Received July 24, 2014

Report

Report Number
1416980-2014-24075
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
June 30, 2014
Report Date
June 30, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IN ORDER TO FURTHER INVESTIGATE THIS CONDITION, A CAPA WAS OPENED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF BAXTER'S INVESTIGATION, IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE LOT WAS MANUFACTURED FROM APRIL 8, 2014 TO APRIL 10, 2014. EVALUATION SUMMARY: VISUAL INSPECTION OF THE RETURNED DEVICE NOTED A SOLID, WHITE PARTICLE, 0.40 MM IN SIZE, FLOATING IN THE FLUID OF THE BLADDER. FOURIER TRANSFORM INFRARED SPECTROSCOPY (FT-IR) SCANNING IDENTIFIED THE PARTICLE TO BE ACRYLIC MATERIAL. THE CAUSE OR ORIGIN OF THE PARTICLE WAS UNABLE TO BE DETERMINED. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

A CUSTOMER REPORTED OBSERVING PARTICULATE MATTER INSIDE THE RESERVOIR OF AN INTERMATE ELASTOMERIC DEVICE. THIS WAS OBSERVED OUTSIDE OF PATIENT USE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
433483 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 14D012

Patients

Seq Age Sex Outcome Treatment
1