FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 3955461 · Received July 24, 2014

Report

Report Number
2531779-2014-21290
Event Type
Malfunction
Date Received
July 24, 2014
Report Date
July 10, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/16/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE ALARM HISTORY INDICATED CONSECUTIVE CALL SERVICE SLEEP ALARMS ON (B)(6) 2014. THE COMPLAINT OF A BLANK DISPLAY COULD NOT BE DUPLICATED ON INVESTIGATION; THE PUMP POWERED ON TO A VISIBLE DISPLAY SCREEN. THE PUMP¿S COVER WAS REMOVED, AND THERE WAS NO INTERMITTENT CONDITION FOUND TO THE DISPLAY FLEX/CONNECTOR. UNRELATED TO THE ORIGINAL COMPLAINT, INVESTIGATION REVEALED THAT THE DISPLAY WAS DIM AND DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
433705 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1