SYMBIQ DUAL CHANNEL
Report
- Report Number
- 9615050-2014-03978
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Date of Event
- March 18, 2014
- Report Date
- May 20, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z0070-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION AT THE SERVICE CENTER FOUND THE DEVICE TOUCHSCREEN WAS RESPONSIVE BUT THE KEYS WERE OUT OF ALIGNMENT AND THEREFORE THE DEVICE COULD NOT BE PROGRAMMED. THE PROBABLE CAUSE WAS DUE TO DRY CONTAMINATION CAUSED BY FLUID INGRESS FOUND ON THE TOUCHSCREEN CONNECTORS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY THE DEVICE TOUCHSCREEN DID NOT PASS THE CALIBRATION TEST. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 346672 | SYMBIQ DUAL CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |