FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 3955450 · Received June 12, 2014

Report

Report Number
9615050-2014-03978
Event Type
Malfunction
Date Received
June 12, 2014
Date of Event
March 18, 2014
Report Date
May 20, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION AT THE SERVICE CENTER FOUND THE DEVICE TOUCHSCREEN WAS RESPONSIVE BUT THE KEYS WERE OUT OF ALIGNMENT AND THEREFORE THE DEVICE COULD NOT BE PROGRAMMED. THE PROBABLE CAUSE WAS DUE TO DRY CONTAMINATION CAUSED BY FLUID INGRESS FOUND ON THE TOUCHSCREEN CONNECTORS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY THE DEVICE TOUCHSCREEN DID NOT PASS THE CALIBRATION TEST. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
346672 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA