FDA Adverse Event
Malfunction
Summary report: N
5.0 CM SHORT ATTACHMENT
MDR report key: 3955396
·
Received December 13, 2013
Report
- Report Number
- 1045834-2013-13551
- Event Type
- Malfunction
- Date Received
- December 13, 2013
- Date of Event
- December 21, 2011
- Report Date
- December 21, 2011
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE REPORTED CONDITION OF "OVERHEATING" WAS CONFIRMED. DURING THE PRE-REPAIR DIAGNOSTIC ASSESSMENT THE DEVICE WAS FOUND TO HAVE DAMAGE TO THE BEARINGS AND DRIED DETERGENT RESIDUE ON INTERNAL SURFACES, INDICATIVE OF IMPROPER CLEANING. IF ADDITIONAL INFO IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).
Description of Event or Problem · 1
A REPORT WAS RECEIVED FROM USA STATING THE DEVICE OVERHEATED DURING SURGERY. NO INJURY OR MED INTERVENTION OCCURRED. THERE WAS NO ADDITIONAL INFO PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 653887 | 5.0 CM SHORT ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|