Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE REPORTED CONDITION OF "OVERHEATING" WAS CONFIRMED. DURING THE PRE-REPAIR DIAGNOSTIC ASSESSMENT THE DEVICE WAS FOUND TO HAVE DAMAGE TO THE BEARINGS AND DRIED DETERGENT RESIDUE ON INTERNAL SURFACES, INDICATIVE OF IMPROPER CLEANING. IF ADDITIONAL INFO IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).

Description of Event or Problem · 1

A REPORT WAS RECEIVED FROM USA STATING THE DEVICE OVERHEATED DURING SURGERY. NO INJURY OR MED INTERVENTION OCCURRED. THERE WAS NO ADDITIONAL INFO PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
653887 5.0 CM SHORT ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment