FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 3955379 · Received June 12, 2014

Report

Report Number
9615050-2014-03973
Event Type
Malfunction
Date Received
June 12, 2014
Date of Event
January 1, 2014
Report Date
May 20, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY, DURING TESTING, THE DEVICE DID NOT ALARM WITH A S233 (OVERTEMPERATURE-PSC) MALFUNCTION ALARM CODE; HOWEVER, IT WAS NOTED IN THE DEVICE HISTORY. THE PROBABLE CAUSE OF THE S233 MALFUNCTION ALARM CODE WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) MALFUNCTION ALARM CODE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, A S233 MALFUNCTION ALARM CODE WAS NOTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347672 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD.

Patients

Seq Age Sex Outcome Treatment
1