FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3955155 · Received July 24, 2014

Report

Report Number
1416980-2014-24060
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
February 15, 2014
Report Date
July 2, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS THE TOTAL TIDAL ULTRA FILTRATION (UF) REMOVAL WAS SET TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 05:25:43. DURING NIGHT DRAIN CYCLE SEVEN, THE PATIENT'S ULTRAFILTRATION READING WAS 2003ML, INDICATING THE HOME PATIENT (HP) DRAINED 1503ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434588 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1