FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3955149 · Received July 24, 2014

Report

Report Number
2531779-2014-21271
Event Type
Malfunction
Date Received
July 24, 2014
Report Date
July 15, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/18/2014 WITH THE FOLLOWING: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED NO ACTIVITY RELATED TO THE COMPLAINT WAS RECORDED. THE BATTERY COMPARTMENT AND BATTERY CAP ARE INTACT WITH NO PHYSICAL DAMAGE OR EVIDENCE OF MOISTURE DAMAGE OBSERVED. DURING TESTING, THE PUMP POWERED ON TO THE ¿VERIFY¿ SCREEN IN ACCORDANCE WITH NORMAL OPERATION. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO OVERHEATING OR ALARMS OCCURRING. THE PUMP¿S ELECTRICAL CURRENT DRAWS WERE FOUND TO BE WITHIN THE REQUIRED SPECIFICATIONS. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF INTERNAL MOISTURE DAMAGE WAS FOUND. THERE WERE NO DEFECTS FOUND TO THE POWER FLEX OR BATTERY CONNECTIONS. THE TEMPERATURE ISSUE WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION. NO DEFECTS WERE FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A TEMPERATURE (TEMP - NO PHYSICAL DAMAGE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434587 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR