FDA Adverse Event Malfunction Summary report: N

GALAXY INTRAVASCULAR ULTRASOUND SYSTEM

MDR report key: 3955132 · Received July 24, 2014

Report

Report Number
2134265-2014-04391
Event Type
Malfunction
Date Received
July 24, 2014
Report Date
June 26, 2014
Manufacturer
BOSTON SCIENTIFIC - FREMONT (SUD)
Product Code
IYO
PMA / PMN Number
K980851
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MFR.: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

SAME CASE AS: 2134265-2014-04386 AND 2134265-2014-04389. IT WAS REPORTED THAT AUTO PULLBACK FAILED. THE ILAB MOTOR DRIVE UNIT WAS USED IN CONJUNCTION WITH A BSC CORONARY IMAGING CATHETER INTENDED TO VISUALIZE THE UNSPECIFIED LESION. DURING PERCUTANEOUS CORONARY INTERVENTION, AN INTRAVASCULAR ULTRASOUND WAS PERFORMED BUT THE MOTOR DRIVE UNIT FAILED TO PERFORM PULLBACK AND COULD ONLY DO A MANUAL PULLBACK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY PERFORMING A MANUAL PULLBACK. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434119 GALAXY INTRAVASCULAR ULTRASOUND SYSTEM SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC IYO BOSTON SCIENTIFIC - FREMONT (SUD) H749A70200

Patients

Seq Age Sex Outcome Treatment
1