ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2014-05189
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Date of Event
- July 11, 2014
- Report Date
- July 14, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS OF THE ER320 DEVICE FOUND THAT IT WAS RECEIVED WITH A CLIP WITHIN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO MEASURE JAWS WIDTH AND THEY WERE FOUND TO BE YIELDED AND MISALIGNED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE CLIPS WERE FED AS INTENDED; HOWEVER, DUE TO THE MISALIGNED CONDITION OF THE JAWS, FIVE SCISSORED CLIPS WERE FORMED. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. AT THE TIME OF THIS SUBMISSION, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS.
IT WAS REPORTED THAT DURING A LAP CHOLECYSTECTOMY, SCISSORING OCCURRED AFTER THE 1ST FIRING IN THE 1ST DEVICE. THE CLIP WAS FED INTO THE JAWS SIDEWAYS FROM THE 1ST FIRING IN THE 2ND DEVICE. THERE WAS NEITHER UNEXPECTED RESISTANCE NOR NOISE IN THE BOTH DEVICES. THERE WAS NO TORQUING OR TWISTING OF THE DEVICE PRESENT. THE DEVICES DID NOT CLAMP SOMETHING HARD SUCH AS AN EXISTING CLIP. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434117 | ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | L4EC12 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |