FDA Adverse Event Malfunction Summary report: N

CAPITAL TEMP PUMP PROFESSIONAL

MDR report key: 3955031 · Received July 24, 2014

Report

Report Number
0001831750-2014-03178
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
June 16, 2014
Report Date
July 22, 2014
Manufacturer
STRYKER MEDICAL-KALAMAZOO
Product Code
ILO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE UNIT WAS REPLACED FOR THE CUSTOMER.

Additional Manufacturer Narrative · 1

UPON FURTHER INVESTIGATION, IT WAS FOUND THAT THE UNIT WOULD NOT HEAT. A NON-FUNCTIONING HEATER WOULD CAUSE THE DEVICE TO NOT DELIVER WATER ACCORDING TO TEMPERATURE SETTING. AN ALTERNATIVE DEVICE COULD BE USED FOR TEMPERATURE THERAPY. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED BY SERVICE REPORT THAT THE UNIT FAILED THE OVERHEAT TEST. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED BY SERVICE REPORT THAT THE UNIT FAILED THE OVERHEAT TEST. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
433996 CAPITAL TEMP PUMP PROFESSIONAL PACK, HOT OR COLD, WATER CIRCULATING ILO STRYKER MEDICAL-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1