CHARGING SYSTEM
Report
- Report Number
- 1627487-2014-02503
- Event Type
- Injury
- Date Received
- July 16, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL;: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKETING HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2014-02502. IT WAS REPORTED THE PT EXPERIENCED HEATING WHILE CHARGING HER SCS IPG. THE PT STATED SHE FELT WARMTH AND THE IPG SITE GETTING HOT WHILE CHARGING. A REPLACEMENT CHARGING SYSTEM WAS SENT TO THE PT TO ADDRESS THE ISSUE. F/U ON (B)(6) 2014 IDENTIFIED THE PT REPORTED THE REPLACEMENT CHARGER RESOLVED THE ISSUE HEATING ISSUE. ON 08/01/2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND THE OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 416101 | CHARGING SYSTEM | SCS CHARGING SYSTEM | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | 3433732 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Other | SCS ANCHOR: MODEL 1192| IMPLANT DATE:| SCS LEAD: MODEL 3228| IMPLANT DATE: |