ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-21248
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Report Date
- July 21, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS EXAMINED AND FOUND THAT THE BATTERY COMPARTMENT WAS CRACKED BETWEEN THE CASE SEAL AND THE GRIP PAD AND CORROSION WAS OBSERVED INSIDE THE BATTERY COMPARTMENT. DURING TESTING, THE PUMP WAS ABLE TO POWER ON SUCCESSFULLY. THE KEYPAD BUTTONS WERE NOTED TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. THE PUMP WAS OPENED AND NO FURTHER MOISTURE WAS OBSERVED INSIDE THE PUMP. UNRELATED TO THE BATTERY COMPARTMENT AND KEYPAD ISSUES, EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. (B)(6).
THE PUMP WAS RETURNED TO ANIMAS. PRODUCT ANALYSIS EVALUATED THE PUMP AND FOUND THAT THE BATTERY COMPARTMENT WAS CRACKED AND THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING TO BUTTON PRESSES. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED (B)(4) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 433407 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |