FDA Adverse Event
Injury
Summary report: N
840 VENTILATOR
MDR report key: 3954627
·
Received July 11, 2014
Report
- Report Number
- 8020893-2014-01710
- Event Type
- Injury
- Date Received
- July 11, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- COVIDIEN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCB'S). THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. (B)(4).
Description of Event or Problem · 1
COVIDIEN RECEIVED INFO STATING THAT WHILE IN USE ON A PT AN 840 VENTILATOR GENERATED AN AUDIBLE AND VISUAL ALARM THEN THE SCREEN WENT BLANK. THE PT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 408008 | 840 VENTILATOR | CONTINUOUS VENTILATOR | CBK | COVIDIEN | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |